Key result
An eight-year-old left ventricular pacemaker lead was successfully extracted percutaneously using an embolic protection system without complications in a patient with a pacemaker pocket infection.
Case Report (n=1)
No
Percutaneous extraction of old, accidentally implanted left ventricular leads is feasible and may be facilitated by cerebral embolic protection and transesophageal echocardiography.
Answers are grounded in this paper and cite their sources.
May support embolic protection during chronic LV lead extraction in pocket infection; hypothesis-generating and requires prospective validation.
Incorrect implantation of a ventricular pacemaker (PM) lead into the left ventricle (LV) is a known problem associated with permanent pacing. The optimal management of such cases identified late has not been clearly established. Generally acceptable management options are: open-chest cardiac surgery using cardio-pulmonary bypass, chronic anticoagulation and antiplatelet-drugs therapy. Rarely, the problem is solved by percutaneous LV lead extraction. We present a case of a patient with DDD pacing and ventricular lead implanted incorrectly into the LV apex region via an atrial septal defect eight years ago. Chronic PM pocket infection developed after replacement of the device. Both leads were extracted percutaneously, and the embolic protection system (Filter-Wire EZ, Boston Scientific) was used to reduce cerebral circulation embolism. The hardest connective tissue adhesions affecting the lead and the anodal ring were found in the LV. Less dense surrounding fibrous tissue around the lead was present at all levels of the venous course of the lead and in the right atrium. Very small fragments of apparently connective tissue remnants were found in cerebral circulation protection filters, and had been removed after the procedure. We conclude that old, permanently implanted LV leads may be extracted percutaneously, especially when there is an increased risk of cardiac surgery, or where the patient's consent for surgical treatment is lacking. In order to perform the procedure it is recommended to establish a cerebral protection system and intraoperative transoesophageal echocardiography which are mandatory for successful lead removal.
No takes yet. Share an insight, caveat, or question.
Kutarski et al. (2013) conducted a case report in Incorrect left ventricular pacemaker lead implantation with pocket infection (n=1). Percutaneous transvenous lead extraction with embolic protection was evaluated on Successful lead extraction. An eight-year-old left ventricular pacemaker lead was successfully extracted percutaneously using an embolic protection system without complications in a patient with a pacemaker pocket infection.