Answers are grounded in this paper and cite their sources.
The areas under the concentration-time curve (AUCs) for Xpert-MTB-HR-Prototype were 0.89 (confidence interval [CI], 0.83 to 0.94) against the CMRS and 0.94 (CI, 0.89 to 0.98) against Xpert MTB/RIF. Considering Xpert-MTB-HR-Prototype as a triage test (at the nearest upper value of sensitivity to 90%), specificities were 55.8% (CI, 47.2 to 64.1%) compared to the CMRS and 85.9% (CI, 79.3 to 90.7%) compared to Xpert MTB/RIF as confirmatory tests. Considering Xpert-MTB-HR-Prototype as a stand-alone diagnostic test, at a specificity near 95%, the test achieved a sensitivity of 65.7% (CI, 53.7 to 75.9%), while the CRP assay achieved a sensitivity of only 13.6% (CI, 7.3 to 23.4%). In this first accuracy study of a prototype blood-based host marker assay, we show the possible value of the assay for triage and diagnosis in PLHIV.
No takes yet. Share an insight, caveat, or question.
Södersten et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: