Why the study?
Cardiac pacing during TAVI and complex PCI generally requires an additional venous access site for a temporary lead, prompting evaluation of the purpose-built Electroducer Sleeve for direct wire pacing.
Does the Electroducer Sleeve device provide safe and effective temporary cardiac pacing in patients undergoing TAVI or PCI?
Does the Electroducer Sleeve device provide safe and effective temporary cardiac pacing in patients undergoing TAVI or PCI?
The Electroducer Sleeve is a safe and effective novel device for temporary cardiac pacing during TAVI and PCI, potentially eliminating the need for a separate temporary venous pacemaker.
Answers are grounded in this paper and cite their sources.
Early data support safety of the Electroducer Sleeve for direct wire pacing; hypothesis-generating and requires randomized validation before adoption.
BACKGROUND: Transcatheter aortic valve implantation (TAVI) and complex percutaneous coronary interventions (PCI) may require cardiac pacing during device delivery, generally requiring the insertion of a temporary pacing lead via an additional venous access site. The purpose-built Electroducer Sleeve device provides direct wire pacing without the need for a temporary venous pacemaker. AIMS: This study assessed the safety of temporary cardiac pacing using the novel sleeve device during PCI. METHODS: This was a multicentre, non-randomised, prospective, first-in-human, single-arm, pilot study. The primary endpoint was analysis of a safety outcome, defined as the occurrence of haematomas or bleeding complications at the device vascular access site. Secondary endpoints included analyses of effectiveness and qualitative outcomes. RESULTS: Sixty patients (mean age: 77.9±9.6 years) from 4 centres in France were included: 39 (65%) underwent TAVI, and 21 (35%) underwent PCI. Procedures were performed using the sleeve with access through the radial (32 patients; 53.3%) or femoral arteries (26; 43.3%), or the femoral vein (2; 3.3%). Primary endpoint analysis revealed that 2 patients (3.3%) developed EArly Discharge After Transradial Stenting of CoronarY Arteries Study (EASY) grade I/Bleeding Academic Research Consortium (BARC) type I haematomas at the device access site. As a measure of effectiveness, a haemodynamic effect was observed after each spike delivery in 54 patients (90%). Analyses of other secondary endpoints showed that 2 patients (6.3%) presented asymptomatic radial artery occlusion. No allergies were reported. CONCLUSIONS: This first-in-human trial using the Electroducer Sleeve indicated that this novel, purpose-built, temporary pacing device was safe and effective. Larger prospective studies are required to confirm these findings.
No takes yet. Share an insight, caveat, or question.
Wintzer-Wehekind et al. (2023) studied this question.