The development of simple sensing devices with rapid analytical readouts is important for practical metabolite analysis, yet many conventional assays rely on enzymes or separate testing formats. Herein, we developed a transparent-film dual-mode sensing platform integrating enzyme-free electrochemical lactate detection with semiquantitative colorimetric cholesterol assessment. The lactate-sensing zone employed a screen-printed graphene electrode modified with palladium nanoparticles (PdNPs) and activated by wide-potential-window cycling. The activated PdNP/graphene interface enhanced charge transfer and enabled lactate detection through a plausible lactate-dependent interfacial process potentially involving modulation of Pd 0 /PdH x -related behavior. A spatially separated ferric chloride/sulfuric acid-based colorimetric zone enabled cholesterol assessment within the same platform while preventing interference between the chemically incompatible assays. Under optimized conditions, the lactate sensor exhibited two linear ranges of 0.1–2 and 2–8 mM, with a detection limit of 82 µM. The cholesterol module provided a linear response from 10 to 500 µM within 1 min, with a detection limit of 10 µM. Analysis of spiked human serum samples yielded recoveries of 97.70–104.01% for lactate and 92.34–104.03% for cholesterol. The proposed platform provides a simple strategy for integrating dual-mode, enzyme-free metabolite analysis.