Answers are grounded in this paper and cite their sources.
We have developed a low-noise digital camera based on a 512 x 96 element CCD operating in the time-delay integration mode. This camera has been combined with an x-ray image intensifier to record radiographic images produced by a scanning slot beam of radiation. This results in the rejection of a large fraction of scattered radiation, without a significant increase in x-ray tube heat loading or image acquisition time. Here we describe the design of our CCD camera and the results of our investigations of camera resolution, linearity, noise, and quantum efficiency. We have found that both the resolution limit (50 mm-1) and the dynamic range (2100) of this novel camera are greater than reported values for conventional video cameras. Applications of this system in digital angiography and mammography are discussed.
No takes yet. Share an insight, caveat, or question.
Holdsworth et al. (1990) studied this question.