Key result
The SUpra-annular PERceval Bentall (SUPERB) procedure was successfully performed in a 75-year-old male, decoupling valve sizing from native annular dimensions to potentially mitigate mismatch.
Why the study?
The authors sought to demonstrate a supra-annular implantation of a composite valve-graft constructed from a sutureless aortic valve within a Dacron graft for aortic root replacement.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
The SUPERB technique demonstrates the feasibility of using a sutureless valve within a Dacron graft for supra-annular aortic root replacement, potentially mitigating patient-prosthetic mismatch.
OBJECTIVE: To demonstrate a supra-annular implantation of a composite valve-graft (CVG) constructed from a sutureless aortic valve within a Dacron graft for aortic root replacement. The anatomic rationale, preclinical validation, and clinical outcomes of this novel technique are described. METHODS: The anatomic rational for sizing of the valve and graft was examined by ex vivo implantation of a Perceval Plus XL valve (Corcym, Milan, Italy) within a 28 mm Valsalva Dacron graft (Terumo Aortic, Inchinnan, UK). Hydrodynamic in vitro testing with a ViVitro pulse duplicator system (ViVitro, Victoria, BC, Canada) validated the integrity of the CVG. The CVG was implanted supra-annular to the native aortic annulus in a clinical setting. RESULTS: A Perceval XL valve within a 28 mm Valsalva Dacron graft provided 8% oversize of the valve inflow ring within the graft collar. In vitro testing showed valve stability within the Dacron graft under physiologic conditions. The first SUpra-annular PERceval Bentall (SUPERB) was performed in a 75-year-old male patient with concomitant aortic stenosis and aortic root aneurysm. The CVG was constructed by (1) suturing the graft to the aortic annulus, (2) construction of coronary buttons, and (3) deployment of the sutureless valve with the Dacron graft in a supra-annular position. CONCLUSIONS: SUPERB uses a Perceval XL valve within a 28 mm Valsalva Dacron graft for aortic root replacement. Positioning of the sutureless valve within the Dacron graft and supra-annular to the native aortic annulus decouples valve sizing from the restrictions of native aortic annular dimensions. This approach simplifies CVG sizing and potentially mitigates patient-prosthetic mismatch.
No takes yet. Share an insight, caveat, or question.
Malaisrie et al. (2026) conducted a case report in Aortic stenosis and aortic root aneurysm (n=1). SUpra-annular PERceval Bentall (SUPERB) procedure was evaluated on Successful implantation and valve stability. The SUpra-annular PERceval Bentall (SUPERB) procedure was successfully performed in a 75-year-old male, decoupling valve sizing from native annular dimensions to potentially mitigate mismatch.
Synapse has enriched 10 related papers. Consider them for comparative context: