DG-PT L Rec liquid reagent showed high correlation with the reference reagent Thromborel S for prothrombin time determination across 581 clinical samples.
Why the study?
Does Grifols' DG-PT L Rec liquid reagent provide comparable prothrombin time determination to Siemens' Thromborel S reference reagent in clinical plasma samples?
Does Grifols' DG-PT L Rec liquid reagent provide comparable prothrombin time determination to Siemens' Thromborel S reference reagent in clinical plasma samples?
The DG-PT L Rec liquid reagent demonstrates high performance and comparability to the reference reagent for prothrombin time determination, supporting its clinical utility.
Answers are grounded in this paper and cite their sources.
Supports liquid reagent use in PT testing; leaves open prospective outcome validation.
The objective is to evaluate Grifols' DG-PT L Rec liquid reagent for prothrombin time (PT) determination in comparison to the laboratory's reference reagent (Siemens' Thromborel S). For linearity, the average master curve for PT and five nominal prothrombin concentrations was obtained from five calibration curves. Within-assay precision (repeatability) was calculated after measuring 20 successive tests of normal and pathological controls. For correlation, 581 routine clinical citrated plasma samples were assessed with both reagents. The BCS XP hemostasis analyzer was used. Linearity of the DG-PT L Rec was good (P < 0.001). The coefficient of variation met the desirable imprecision of less than 2% (normal controls: 1.7%; pathological controls: 0.9%). Correlation between DG-PT L Rec and Thromborel S was high (r = 0.9795; PT in %). In subgroups of anticoagulated, low fibrinogen, lipemic, jaundice, and hemolyzed samples the correlation was more than 0.95. Performance of DG-PT L Rec was high and comparable to the reference reagent.
No takes yet. Share an insight, caveat, or question.
Segarra et al. (2016) studied this question. DG-PT L Rec liquid reagent vs. Siemens' Thromborel S was evaluated on Correlation of prothrombin time determination (r = 0.9795). DG-PT L Rec liquid reagent showed high correlation with the reference reagent Thromborel S for prothrombin time determination across 581 clinical samples.