Key result
Remote-controlled magnetic mapping and ablation of right ventricular outflow tract tachycardia was successful in 92.3% of patients, with no complications and rare mechanically induced extrasystoles.
Why the study?
Does remote-controlled magnetic mapping and ablation achieve successful ablation in patients with right ventricular outflow tract tachycardia?
Observational (n=13)
Does remote-controlled magnetic mapping and ablation achieve successful ablation in patients with right ventricular outflow tract tachycardia?
Remote-controlled magnetic mapping and ablation of RVOT tachycardia is feasible, safe, and effective, with a high success rate and low incidence of mechanically induced extrasystoles.
Answers are grounded in this paper and cite their sources.
Supports feasibility in experienced centers; leaves open need for randomized trials versus manual ablation.
AIMS: Manual mapping and ablation of right ventricular outflow tract (RVOT) tachycardia may be associated with cardiac perforation and complicated by mechanically induced ventricular extrasystoles (VESs). The aim of this study was to assess remote-controlled magnetic (RCM) mapping and ablation of RVOT ventricular tachycardia (VT)/VES utilizing a soft magnetic catheter. METHODS AND RESULTS: Mapping and ablation of RVOT VT/VES were performed using the magnetic navigation system (MNS, Niobe II, Stereotaxis, St Louis, MO) in conjunction with a cardiodrive motor unit (Stereotaxis). A soft magnetic catheter (Celsius RMT, 4 mm solid tip, Biosense Webster, Diamond Bar, CA) was navigated to the RVOT using a sequence of pre-installed magnetic vectors. The primary endpoint was defined as successful RCM VT/VES ablation. Other parameters observed included RCM RVOT accessibility, mapping-induced mechanical VES, fluoroscopy time, complications, and arrhythmia recurrence. Thirteen consecutive patients (mean age: 50 ± 14 years; 10 females, 6 RVOT VT) underwent RCM mapping and ablation. The RVOT was reached in all patients utilizing solely the MNS. Successful RCM RVOT ablation was achieved in {12 of 13} (92.3%) patients. Mean procedure and fluoroscopy times were 116.5 ± 48.9 and 7.5 ± 4.3 min, respectively. Catheter-induced VES during RVOT mapping were observed in {1 in 13} patients (7.7%, three VES). No complications occurred. During a mean follow-up of 252 ± 211 days, clinical arrhythmia recurrence was observed in {1 in 13} (7.7%) patients. CONCLUSIONS: Automatic RCM RVOT access is feasible, while RVOT mapping and ablation appear to be safe, fast, and effective. The soft magnetic catheter rarely induces mechanical VES.
No takes yet. Share an insight, caveat, or question.
Konstantinidou et al. (2011) conducted an observational in Right ventricular outflow tract (RVOT) tachycardia and ventricular extrasystoles (n=13). Remote-controlled magnetic (RCM) mapping and ablation was evaluated on Successful RCM VT/VES ablation. Remote-controlled magnetic mapping and ablation of right ventricular outflow tract tachycardia was successful in 92.3% of patients, with no complications and rare mechanically induced extrasystoles.
Synapse has enriched 10 related papers. Consider them for comparative context: