Answers are grounded in this paper and cite their sources.
PURPOSE: Amide proton transfer (APT) imaging may detect changes in tissues' pH based on the chemical exchange saturation transfer (CEST) phenomenon, and thus it may be useful for identifying the penumbra in ischemic stroke patients. We investigated the effect of saturation pulse duration and power on the APT effect in phantoms with different pH values. METHODS: . The dependence of the APT SI on pH and the relative change in the APT SI were calculated as the slope of the linear regression. RESULTS: The lower the pH, the larger the relative change in the APT SI, due to the change in saturation pulse duration and power. The APT SI was highly correlated with the pH at all saturation pulse durations and powers. CONCLUSION: The influence of saturation duration and power on the APT effect was greater at lower pH than higher pH. The combination of saturation pulse ≥ 1.0 s and power ≥ 1.5 μT was useful for the sensitive detection of changes in APT effects in the egg-white albumin samples with different pH values.