Key result
InfoBionic MoMe monitor captures ~78% of standard telemetry alarms, detecting all major arrhythmias.
Why the study?
To investigate the operational outcomes and implementation effects of tiered cardiac telemetry monitoring in a hospital environment using an innovative technology.
Does a cellular-connected wireless continuous cardiac monitor (InfoBionic MoMe) provide comparable alert precision, speed, and reliability to standard hardwired cardiac telemetry in adult inpatients with acute cardiac conditions?
Observational (n=162)
Open-label
No
Does a cellular-connected wireless continuous cardiac monitor (InfoBionic MoMe) provide comparable alert precision, speed, and reliability to standard hardwired cardiac telemetry in adult inpatients with acute cardiac conditions?
While a cellular-connected wireless telemetry device achieved 100% capture for major arrhythmias like ventricular tachycardia and atrial fibrillation, significant alarm delays from cellular connectivity issues and 22% missed alarms require technological and infrastructure improvements before inpatient clinical implementation.
Answers are grounded in this paper and cite their sources.
Should not yet change inpatient telemetry practice due to missed alarms and delays; leaves open feasibility after technological refinements.
Objective: To investigate the operational outcomes and implementation effects of tiered cardiac telemetry monitoring in a hospital environment using an innovative technology. Patients and Methods: The research focuses on assessing the precision, speed, and reliability of alerts generated by a wireless device in adult patients aged 18 and above, concurrently monitored by a hardwired, continuous cardiac monitor. Using an agile methodology, we tested and validated a nonhardwired, cellular-connected continuous cardiac monitor (InfoBionic MoMe) in 162 patients. A comparison was made between the wireless device and the standard hardwired system, conducted at Mayo Clinic Hospital with Institutional Review Board approval from June 6, 2022, to December 15, 2022. Results: The study revealed a high correlation of events captured compared with the standard care model. Differences in algorithms, alarm parameters, and operational considerations impacting clinical implementation were observed. Connectivity improvements during the study reduced latency from 3-5 minutes to 30 seconds. Delayed alarms were attributed to device damage (4.5% of cases) and poor cellular connections (29% within 31-60 seconds). Conclusion: The implementation of tiered cardiac telemetry in hospital environments, coupled with advancements in remote cardiac monitoring, supports expanded bedside telemetry capabilities and near real-time remote monitoring postdischarge. Although the study successfully validated the wireless device concept, improvements are needed before implementation for inpatient cardiac monitoring. Further research and technological enhancements can build on these findings to enhance health care practices in this domain.
No takes yet. Share an insight, caveat, or question.
Disrud et al. (2024) conducted an observational in Acute cardiac conditions requiring continuous monitoring for potential arrhythmias (n=162). InfoBionic MoMe wireless cellular-connected continuous cardiac monitor vs. Standard hardwired inpatient continuous cardiac monitor was evaluated on Capture rate of standard telemetry alarms by wireless device. The InfoBionic MoMe wireless monitor captured 78% of standard system alarms with an average delay of 27 seconds and 100% detection of major arrhythmias.