Why the study?
Previous studies found different cutoffs of vena contracta area to define severe TR, prompting investigation into factors associated with this variability by comparing technical variables and vendors.
Does the choice of echocardiography vendor and technical settings affect the measurement of vena contracta area for grading tricuspid regurgitation severity in patients scheduled for tricuspid surgery?
Does the choice of echocardiography vendor and technical settings affect the measurement of vena contracta area for grading tricuspid regurgitation severity in patients scheduled for tricuspid surgery?
Vena contracta area measurement for tricuspid regurgitation is highly sensitive to technical settings and echocardiography vendor, necessitating consistent methodology for longitudinal analysis.
VCA cutoffs for severe TR vary by vendor and technique; leaves open need for standardized protocols before routine use.
BACKGROUND: Previous studies found different cutoffs of vena contracta area (VCA) to define severe tricuspid regurgitation (TR). The aim of this study was to investigate the factors associated with such variability by comparing technical variables and vendors. METHODS: Sixty-nine patients with scheduled tricuspid surgery were included in this prospective study. For each patient, TR data sets were obtained on three-dimensional color Doppler transthoracic echocardiography on at least two of three systems: GE Vivid E95 (n = 39), Siemens SC2000 Prime (n = 64), and Philips EPIQ 7C (n = 60). VCA was measured using default settings or with color baseline shifted on all three platforms and with minimal color gain (10%-20%) on the GE platform. RESULTS: Color gain reduction and baseline shift caused significant change sin VCA measurement (-46% and 10%, respectively). Intervendor comparison exhibited wide limits of agreement (narrowest range, -74% to 167%), with either default or optimized settings. Different technical settings, platforms, and reference methods all produced different VCA cutoffs for severe TR. CONCLUSIONS: VCA measurement in TR is sensitive to technical factors and demonstrates intervendor variability. Technical variables in VCA measurement should be reported in detail to allow comparison among research studies. The same vendor and settings should be used for longitudinal analysis of TR VCA in the same patient in multivendor echocardiography laboratories.