Key result
The Minicare point-of-care cTnI test demonstrated high diagnostic accuracy for acute myocardial infarction at 3 hours, with an ROC AUC of 0.93, sensitivity of 0.90, and specificity of 0.96.
Why the study?
Does a point-of-care cardiac troponin-I test accurately diagnose acute myocardial infarction in adult patients with suspected acute coronary syndrome?
Observational (n=465)
Yes
Does a point-of-care cardiac troponin-I test accurately diagnose acute myocardial infarction in adult patients with suspected acute coronary syndrome?
Answers are grounded in this paper and cite their sources.
The point-of-care cTnI test demonstrates high diagnostic accuracy for acute myocardial infarction within a 0/3-hour sampling protocol, supporting its clinical utility in suspected acute coronary syndrome.
BACKGROUND: We evaluated the clinical performance of the Minicare cardiac troponin-I (cTnI), a new point-of-care (POC) cTnI test for the diagnosis of acute myocardial infarction (AMI) in a prospective, multicentre study (ISRCTN77371338). METHODS: Of 474 patients (≥18 years) admitted to an emergency department (ED) or chest pain unit (CPU) with symptoms suggestive of acute coronary syndrome (ACS; ≤12 h from symptom onset), 465 were eligible. Minicare cTnI was tested immediately, 3 h and 6 h after presentation. AMI diagnoses were adjudicated independently based on current guidelines. RESULTS: The diagnostic performance of the Minicare cTnI test at 3 h was similar for whole blood and in plasma: sensitivity 0.92 vs. 0.90; specificity 0.91 vs. 0.90; positive predictive value (PPV) 0.68 vs. 0.66; negative predictive value (NPV) 0.98 vs. 0.98; positive likelihood ratio (LR+) 10.18 vs. 9.41; negative likelihood ratio (LR-) 0.09 vs. 0.11. The optimal diagnostic performance was obtained at 3 h using cut-offs cTnI >43 ng/L plus cTnI change from admission ≥18.5 ng/L: sensitivity 0.90, specificity 0.96, PPV 0.81, NPV 0.98, and LR+ 21.54. The area under the receiver operating characteristics (ROC) curve for cTnI whole blood baseline value and absolute change after 3 h curve was 0.93. CONCLUSIONS: These data support the clinical usefulness of Minicare cTnI within a 0 h/3 h-blood sampling protocol supported by current guidelines for the evaluation of suspected ACS.
No takes yet. Share an insight, caveat, or question.
Christ et al. (2018) conducted an observational in acute coronary syndrome (n=465). Minicare cardiac troponin-I (cTnI) test was evaluated on Diagnosis of acute myocardial infarction. The Minicare point-of-care cTnI test demonstrated high diagnostic accuracy for acute myocardial infarction at 3 hours, with an ROC AUC of 0.93, sensitivity of 0.90, and specificity of 0.96.