Why the study?
Does a novel point-of-care cTnI assay provide comparable diagnostic accuracy to current point-of-care and high-sensitivity central laboratory cTnI assays for ruling out acute myocardial infarction in emergency department patients?
Does a novel point-of-care cTnI assay provide comparable diagnostic accuracy to current point-of-care and high-sensitivity central laboratory cTnI assays for ruling out acute myocardial infarction in emergency department patients?
The novel Minicare point-of-care cTnI assay demonstrates comparable diagnostic performance to central laboratory high-sensitivity cTnI for the early rule-out of acute myocardial infarction.
Answers are grounded in this paper and cite their sources.
Supports the clinical utility of this novel point-of-care assay for rapid myocardial infarction.
BACKGROUND: Efficient rule-out of acute myocardial infarction (MI) facilitates early disposition of chest pain patients in emergency departments (ED). Point-of-care (POC) cardiac troponin (cTn) may improve patient throughput. We compared the diagnostic accuracy of a novel cTnI test (Minicare cTnI, Philips), with current POC cTnI (I-Stat, Abbott) and high-sensitivity central laboratory cTnI (hs-cTnI; Architect, Abbott) assays. METHODS: The clinical performance of the assays were compared in samples from 450 patients from a previous clinical evaluation of Minicare cTnI. Areas under the receiver operating characteristics curves were 0.87-0.91 at admission (p=ns) and 0.96-0.97 3h after admission (p=ns). The negative predictive values (NPV) at admission were 95% ((92-97%, 95% CI) for Minicare cTnI and increased to 99% (97-100%) at 2-4h, and similar to Architect hs-cTnI (98%, 96-100%), but higher than I-Stat cTnI (95%, 92-97%; p<0.01). Negative likelihood ratios (LR-) after 2-4h were 0.06 (0.02-0.17, 95% CI) for Minicare cTnI, 0.11 (0.05-0.24) for Architect hs-cTnI (p=0.02) and 0.28 (0.18-0.43) for I-Stat cTnI (p<0.0001). The clinical concordances between Minicare cTnI and Architect hs-cTnI were 92% (admission) and 95% (2-4h), with lower concordances between Minicare cTnI and I-Stat cTnI (83% and 78%, respectively; p=0.007). CONCLUSIONS: The Minicare cTnI POC assay may become useful for prompt and safe ruling-out of AMI in ED patients with suspected AMI using a guideline supported 0/3h sampling protocol.
No takes yet. Share an insight, caveat, or question.
Venge et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: