Key result
The novel in-airway laser absorption spectroscopy device provided highly repeatable measurements of cardiac output, demonstrating an overall coefficient of variation of 6.9% across 5 consecutive days.
Why the study?
Existing limitations in gas analysis needed to be addressed to enable highly accurate, respiratory-based cardiac output measurements rooted in mass balance.
Observational (n=6)
No
A novel in-airway laser absorption spectroscopy device using acetylene and methane tracer gases provides highly precise, non-invasive measurements of cardiac output at rest and during exercise.
Answers are grounded in this paper and cite their sources.
May enable frequent noninvasive cardiac output monitoring; leaves open prospective clinical validation.
Respiratory approaches to determining cardiac output in humans are securely rooted in mass balance and therefore potentially highly accurate. To address existing limitations in the gas analysis, we developed an in-airway analyser based on laser absorption spectroscopy to provide analyses every 10 ms. The technique for estimating cardiac output requires both a relatively soluble and insoluble tracer gas, and we employed acetylene and methane for these, respectively. A multipass cell was used to provide sufficient measurement sensitivity to enable analysis directly within the main gas stream, thus avoiding errors introduced by sidestream gas analysis. To assess performance, measurements of cardiac output were made during both rest and exercise on five successive days in each of six volunteers. The measurements were extremely repeatable (coefficient of variation ~ 7%). This new measurement technology provides a stable foundation against which the algorithm to calculate cardiac output can be further developed.
No takes yet. Share an insight, caveat, or question.
Smith et al. (2021) conducted an observational in Healthy volunteers (n=6). In-airway laser absorption spectroscopy (5-gas MFS) was evaluated on Coefficient of variation for cardiac output measurements. The novel in-airway laser absorption spectroscopy device provided highly repeatable measurements of cardiac output, demonstrating an overall coefficient of variation of 6.9% across 5 consecutive days.