Key result
Non-invasive cardiac output measurement using carbon dioxide partial rebreathing consistently underestimated cardiac output compared to thermodilution, with a bias of -0.60 +/- 0.87 l/min.
Why the study?
Does a non-invasive carbon dioxide partial rebreathing technique accurately measure cardiac output compared to bolus pulmonary artery thermodilution in postoperative cardiac surgery patients?
Observational (n=24)
No
Does a non-invasive carbon dioxide partial rebreathing technique accurately measure cardiac output compared to bolus pulmonary artery thermodilution in postoperative cardiac surgery patients?
Mean Difference: -0.6
The non-invasive carbon dioxide partial rebreathing technique provides reliable but consistently underestimated cardiac output measurements compared to standard thermodilution in postoperative cardiac patients.
Caution advised against sole reliance on non-invasive CO2 rebreathing for CO monitoring post-cardiac surgery; leaves open its clinical role pending larger validation.
This study investigated the accuracy of a new technique for measuring cardiac output using the derivative Fick principle based on the ratio of change in the partial pressures of end-tidal and mixed expired carbon dioxide produced by short periods of partial rebreathing. A prospective clinical study involving 24 patients following cardiopulmonary bypass for coronary artery bypass grafting or valvular surgery was undertaken in the intensive care unit of a university-affiliated hospital. Haemodynamic measurements were performed after admission to the intensive care unit. Cardiac output was measured simultaneously by bolus pulmonary artery thermodilution and by a non-invasive carbon dioxide partial rebreathing technique. Cardiac output measurement using the new technique demonstrated a significant but consistent underestimate, with a bias of -0.60 +/- 0.87 l/min. This new adaptation of the partial rebreathing technique is reliable in measuring cardiac output in postoperative patients. Reasons for the consistent discrepancy between thermodilution and partial rebreathing techniques are discussed.
No takes yet. Share an insight, caveat, or question.
Killick et al. (2008) conducted an observational in Postoperative cardiac surgery (n=24). Non-invasive carbon dioxide partial rebreathing technique vs. Bolus pulmonary artery thermodilution was evaluated on Cardiac output measurement bias (bias of -0.60 +/- 0.87 l/min). Non-invasive cardiac output measurement using carbon dioxide partial rebreathing consistently underestimated cardiac output compared to thermodilution, with a bias of -0.60 +/- 0.87 l/min.