Does freehand dynamic 3D colour flow imaging reveal non-uniform mitral blood flow velocity profiles in healthy volunteers?
Does freehand dynamic 3D colour flow imaging reveal non-uniform mitral blood flow velocity profiles in healthy volunteers?
Demonstrates that mitral blood flow velocity profiles are non-uniform, highlighting potential errors of 10-60% in stroke volume calculations when using a single sample volume in Doppler measurements.
May bias single-point Doppler stroke volume estimates; leaves open whether 3D imaging improves accuracy in practice.
AIMS: To describe the mitral blood flow velocity distribution, we applied a freehand dynamic three-dimensional (3D) colour flow method using a moving sample surface that followed the mitral apparatus during diastole. METHODS: Nineteen healthy volunteers were studied. The ultrasound data were captured from 10-20 heartbeats at high frame rate (mean 46 frames/s) while freely tilting the transducer in an apical position. A magnetic position sensor system recorded the spatial position and orientation of the probe. Blood flow velocities were integrated across a spherical surface. In volumetric blood flow measurements this would yield angle independence of the Doppler beam. Raw digital data were analysed off-line with no loss of temporal resolution. RESULTS: The ratio of the maximum velocity time integral (VTI) to the mean VTI was mean 1.3 (range 1.1-1.6). At the time of peak flow the ratio of the maximum to the mean velocity was mean 1.5 (range 1.2-2.6). CONCLUSION: The blood flow velocity profile was non-uniform. By using a single sample volume in Doppler measurements of the maximum VTI errors ranging from 10 to 60% may be introduced in calculations of stroke volumes.
No takes yet. Share an insight, caveat, or question.
Bjørn Olav Haugen (2000) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: