Key result
Automated 4D flow CMR matches Doppler echocardiography for measuring peak E-wave mitral inflow velocity.
Why the study?
To validate 4D flow CMR peak velocity tracking methods for measuring the peak velocity of mitral inflow against Doppler echocardiography.
Does an automated dynamic peak velocity tracking method using 4D flow CMR accurately measure mitral inflow velocities compared to Doppler echocardiography in patients?
Observational (n=50)
Yes
Does an automated dynamic peak velocity tracking method using 4D flow CMR accurately measure mitral inflow velocities compared to Doppler echocardiography in patients?
Absolute Event Rate: 0.94% vs 0.9%
p-value: p=0.29
Automated dynamic peak velocity tracing using 4D flow CMR is comparable to Doppler echocardiography for mitral inflow assessment.
Supports automated 4D flow CMR mitral inflow assessment; leaves open prospective validation before clinical adoption.
OBJECTIVE: We aim to validate four-dimensional flow cardiovascular magnetic resonance (4D flow CMR) peak velocity tracking methods for measuring the peak velocity of mitral inflow against Doppler echocardiography. METHOD: Fifty patients were recruited who had 4D flow CMR and Doppler Echocardiography. After transvalvular flow segmentation using established valve tracking methods, peak velocity was automatically derived using three-dimensional streamlines of transvalvular flow. In addition, a static-planar method was used at the tip of mitral valve to mimic Doppler technique. RESULTS: Peak E-wave mitral inflow velocity was comparable between TTE and the novel 4D flow automated dynamic method (0.9 ± 0.5 vs 0.94 ± 0.6 m/s; p = 0.29) however there was a statistically significant difference when compared with the static planar method (0.85 ± 0.5 m/s; p = 0.01). Median A-wave peak velocity was also comparable across TTE and the automated dynamic streamline (0.77 ± 0.4 vs 0.76 ± 0.4 m/s; p = 0.77). A significant difference was seen with the static planar method (0.68 ± 0.5 m/s; p = 0.04). E/A ratio was comparable between TTE and both the automated dynamic and static planar method (1.1 ± 0.7 vs 1.15 ± 0.5 m/s; p = 0.74 and 1.15 ± 0.5 m/s; p = 0.5 respectively). Both novel 4D flow methods showed good correlation with TTE for E-wave (dynamic method; r = 0.70; P < 0.001 and static-planar method; r = 0.67; P < 0.001) and A-wave velocity measurements (dynamic method; r = 0.83; P < 0.001 and static method; r = 0.71; P < 0.001). The automated dynamic method demonstrated excellent intra/inter-observer reproducibility for all parameters. CONCLUSION: Automated dynamic peak velocity tracing method using 4D flow CMR is comparable to Doppler echocardiography for mitral inflow assessment and has excellent reproducibility for clinical use.
No takes yet. Share an insight, caveat, or question.
Njoku et al. (2022) reported an observational. Automated dynamic peak velocity tracking using 4D flow CMR vs. Doppler echocardiography (TTE) was evaluated on Peak E-wave mitral inflow velocity (p=0.29). Automated dynamic peak velocity tracking using 4D flow CMR was comparable to Doppler echocardiography for measuring peak E-wave mitral inflow velocity (0.94 vs 0.9 m/s; p=0.29).