Key result
A novel 2.4-GHz frequency-locked loop radar accurately measured respiration and heart rates during nonperiodic body movement, with standard deviations of 1.27 and 5.69 beats/min respectively while seated.
Why the study?
Most conventional vital-sign radars cannot easily detect human vital sign signals under nonperiodic body movement.
A novel 2.4-GHz FLL radar can accurately detect human respiration and heart rates even in the presence of nonperiodic body movements.
Answers are grounded in this paper and cite their sources.
May enable motion-robust non-contact vital sign monitoring; leaves open clinical validation before practice adoption.
Most conventional vital-sign radars cannot easily detect human vital sign signals under nonperiodic body movement. This article presents a novel frequency-locked loop (FLL) radar that can track the phase changes that are caused by nonperiodic body movement. A gain/phase detector and a cross correlation function (CCF) in the FLL radar greatly reduce the second harmonic, the intermodulation terms between vital sign signals and nonperiodic body movement noise, enhancing both the signal-to-noise ratio and the accuracy of detection of the vital signs. Measurements indicated that the implemented 2.4-GHz FLL radar with a gain/phase detector and CCF can effectively detect the human respiration rate (RR) and heart rate (HR) under nonperiodic body movement. Using a chest belt as the measurement reference, the standard deviation (SD) of RR and HR measurements were 1.27 and 5.69 beats/min, respectively, when the seated subject under test (SUT) swung nonperiodically his upper body. When the SUT walked away from the radar naturally at 2.6 km/h, the implemented FLL radar measured RR and HR with SDs of 2.67 and 8.28 beats/min, respectively.
No takes yet. Share an insight, caveat, or question.
Peng et al. (2021) studied this question. 2.4-GHz frequency-locked loop (FLL) radar vs. Chest belt was evaluated on Standard deviation (SD) of respiration rate (RR) and heart rate (HR) measurements. A novel 2.4-GHz frequency-locked loop radar accurately measured respiration and heart rates during nonperiodic body movement, with standard deviations of 1.27 and 5.69 beats/min respectively while seated.