Does sample matrix and storage condition affect BNP measurement accuracy on the Bayer Advia Centaur?
Does sample matrix and storage condition affect BNP measurement accuracy on the Bayer Advia Centaur?
K2 EDTA plasma is the most suitable specimen for BNP testing on the Bayer Advia Centaur for both fresh and frozen samples stored up to 1 week.
May support K2 EDTA preference for BNP on Advia Centaur; leaves open platform-specific matrix validation needs.
The assessment and management of congestive heart failure relies increasingly on the measurement of B-type natriuretic peptide (BNP). However, the effective contribution of this biochemical test in the clinical decision making is influenced by reliability of the measure, which also depends on several preanalytical issues. Since there is controversy on the influence of the matrix and the storage conditions on BNP measurement, we compared results of BNP in serum, K2 ethylene diamine tetra-acetic acid (EDTA) plasma and lithium heparin plasma fresh samples and in matching samples stored at -20 and -80 degrees C for 1 week. BNP measured on the Bayer Advia Centaur was systematically underestimated in heparin plasma (-47%) and serum (-62%) when compared to K2 EDTA plasma. According to the established 100 ng/L cutoff value, 25% and 37% of the fresh samples collected in heparin plasma or serum were misclassified from the reference K2 EDTA fresh specimen, respectively. When compared to the fresh specimens, the mean and interindividual bias observed for samples stored at either -20 degrees C or -80 degrees C was, overall, modest for K2 EDTA plasma (-2%) and heparin plasma (+6% and -4%, respectively), though it appeared clinically meaningful in serum (+47% and +28%, respectively). Although we can not rule out that other BNP assays using different antibodies may be not affected from degradation during storage to the same extent, results of our investigation demonstrate that K2 EDTA plasma is the most suitable specimens for BNP testing on fresh and frozen samples stored at either -20 degrees C or -80 degrees C for up to 1 week.
No takes yet. Share an insight, caveat, or question.
Lippi et al. (2007) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: