Why the study?
Recent evidence suggested higher ICD lead failure in multi-lead defibrillator therapy via the cephalic route, motivating an evaluation of CRT-D implantation via the cephalic vein and ICD lead failure.
Does ICD lead implantation via the cephalic vein prevent ICD lead failure in patients receiving CRT-D?
Does ICD lead implantation via the cephalic vein prevent ICD lead failure in patients receiving CRT-D?
CRT-D implantation via the cephalic route is not associated with an increased risk of premature ICD lead failure compared to non-cephalic routes, though female gender and higher BMI are independent predictors of failure.
Cephalic CRT-D access remains viable without excess lead failure risk; leaves open whether female sex or BMI warrant targeted lead monitoring.
BACKGROUND: Cardiac resynchronization therapy-defibrillator (CRT-D) implantation via the cephalic vein is feasible and safe. Recent evidence has suggested a higher implantable cardioverter-defibrillator (ICD) lead failure in multi-lead defibrillator therapy via the cephalic route. We evaluated the relationship between CRT-D implantation via the cephalic and ICD lead failure. METHODS: Data was collected from three CRT-D implanting centers between October 2008 and September 2017. In total 633 patients were included. Patient and lead characteristics with ICD lead failure were recorded. Comparison of "cephalic" (ICD lead via cephalic) versus "non-cephalic" (ICD lead via non-cephalic route) cohorts was performed. Kaplan-Meier survival and a Cox-regression analysis were applied to assess variables associated with lead failure. RESULTS: The cephalic and non-cephalic cohorts were equally male (81.9% vs. 78%; p = .26), similar in age (69.7 ± 11.5 vs. 68.7 ± 11.9; p = .33) and body mass index (BMI) (27.7 ± 5.1 vs. 27.1 ± 5.7; p = .33). Most ICD leads were implanted via the cephalic vein (73.5%) and patients had a mean of 2.9 ± 0.28 leads implanted via this route. The rate of ICD lead failure was low and statistically similar between both groups (0.36%/year vs. 0.13%/year; p = .12). Female gender was more common in the lead failure cohort than non-failure (55.6% vs. 17.9%, respectively; p = .004) as was hypertension (88.9% vs. 54.2%, respectively, p = .038). On multivariate Cox-regression, female sex (p = .008; HR, 7.12 [1.7-30.2]), and BMI (p = .047; HR, 1.12 [1.001-1.24]) were significantly associated with ICD lead failure. CONCLUSION: CRT-D implantation via the cephalic route is not significantly associated with premature ICD lead failure. Female gender and BMI are predictors of lead failure.
No takes yet. Share an insight, caveat, or question.
Akhtar et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: