Why the study?
Dysfunctional defibrillator leads are commonly abandoned when a new lead is implanted, but long-term follow-up data on abandoned leads remain sparse.
What is the long-term incidence of extraction and associated complications for abandoned transvenous defibrillator leads?
What is the long-term incidence of extraction and associated complications for abandoned transvenous defibrillator leads?
More than 90% of abandoned defibrillator leads do not require extraction during long-term follow-up, and when needed, extraction can be performed with high success and low complication rates.
Abandoned defibrillator leads warrant vigilant follow-up for extraction needs; leaves open optimal management strategies pending randomized data.
AIMS: Commonly, a dysfunctional defibrillator lead is abandoned and a new lead is implanted. Long-term follow-up data on abandoned leads are sparse. We aimed to investigate the incidence and reasons for extraction of abandoned defibrillator leads in a nationwide cohort and to describe extraction procedure-related complications. METHODS AND RESULTS: All abandoned transvenous defibrillator leads were identified in the Danish Pacemaker and ICD Register from 1991 to 2019. The event-free survival of abandoned defibrillator leads was studied, and medical records of patients with interventions on abandoned defibrillator leads were audited for procedure-related data. We identified 740 abandoned defibrillator leads. Meantime from implantation to abandonment was 7.2 ± 3.8 years with mean patient age at abandonment of 66.5 ± 13.7 years. During a mean follow-up after abandonment of 4.4 ± 3.1 years, 65 (8.8%) abandoned defibrillator leads were extracted. Most frequent reason for extraction was infection (pocket and systemic) in 41 (63%) patients. Procedural outcome after lead extraction was clinical success in 63 (97%) patients. Minor complications occurred in 3 (5%) patients, and major complications in 1 (2%) patient. No patient died from complication to the procedure during 30-day follow-up after extraction. CONCLUSION: More than 90% of abandoned defibrillator leads do not need to be extracted during long-term follow-up. The most common indication for extraction is infection. Abandoned defibrillator leads can be extracted with high clinical success rate and low risk of major complications at high-volume centres.
No takes yet. Share an insight, caveat, or question.
Elgaard et al. (2020) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: