Key result
Optical diffuse correlation spectroscopy demonstrated good agreement with transcranial Doppler ultrasound for measuring critical closing pressure in healthy adults (R = 0.66).
Why the study?
Does optical diffuse correlation spectroscopy (DCS) accurately measure critical closing pressure compared to transcranial Doppler ultrasound in healthy adults?
Observational (n=18)
Does optical diffuse correlation spectroscopy (DCS) accurately measure critical closing pressure compared to transcranial Doppler ultrasound in healthy adults?
Effect estimate: R = 0.66
Absolute Event Rate: 11.1% vs 13%
Optical diffuse correlation spectroscopy provides a non-invasive, continuous method for monitoring critical closing pressure and arteriole compliance that agrees well with transcranial Doppler ultrasound.
May support noninvasive cerebral pressure monitoring; leaves open validation in clinical populations.
The critical closing pressure ( CrCP) of the cerebral circulation depends on both tissue intracranial pressure and vasomotor tone. CrCP defines the arterial blood pressure ( ABP) at which cerebral blood flow approaches zero, and their difference ( ABP - CrCP) is an accurate estimate of cerebral perfusion pressure. Here we demonstrate a novel non-invasive technique for continuous monitoring of CrCP at the bedside. The methodology combines optical diffuse correlation spectroscopy (DCS) measurements of pulsatile cerebral blood flow in arterioles with concurrent ABP data during the cardiac cycle. Together, the two waveforms permit calculation of CrCP via the two-compartment Windkessel model for flow in the cerebral arterioles. Measurements of CrCP by optics (DCS) and transcranial Doppler ultrasound (TCD) were carried out in 18 healthy adults; they demonstrated good agreement (R = 0.66, slope = 1.14 ± 0.23) with means of 11.1 ± 5.0 and 13.0 ± 7.5 mmHg, respectively. Additionally, a potentially useful and rarely measured arteriole compliance parameter was derived from the phase difference between ABP and DCS arteriole blood flow waveforms. The measurements provide evidence that DCS signals originate predominantly from arteriole blood flow and are well suited for long-term continuous monitoring of CrCP and assessment of arteriole compliance in the clinic.
No takes yet. Share an insight, caveat, or question.
Baker et al. (2017) conducted an observational in Healthy (n=18). Optical diffuse correlation spectroscopy (DCS) vs. Transcranial Doppler ultrasound (TCD) was evaluated on Critical closing pressure (CrCP) (R = 0.66). Optical diffuse correlation spectroscopy demonstrated good agreement with transcranial Doppler ultrasound for measuring critical closing pressure in healthy adults (R = 0.66).