Key result
Entrainment mapping identifies far-field potentials at ~86% of double potential sites, improving VT mapping accuracy.
Why the study?
Electrograms with multiple peaks commonly complicate mapping and identification of catheter ablation targets in infarcts.
Does entrainment mapping distinguish far-field potentials from local potentials during mapping of ventricular tachycardia in patients with prior infarction?
Observational (n=13)
Does entrainment mapping distinguish far-field potentials from local potentials during mapping of ventricular tachycardia in patients with prior infarction?
Answers are grounded in this paper and cite their sources.
Entrainment mapping can effectively distinguish far-field from local potentials in infarct-related ventricular tachycardia, improving mapping accuracy and preventing erroneous site classification.
OBJECTIVES: The goal of this study was to assess entrainment for distinguishing far-field potentials (FFP) due to depolarization of tissue at a distance from the mapping catheter from the local potential (LP) due to depolarization of tissue at the catheter electrode during mapping of ventricular tachycardia (VT). BACKGROUND: Electrograms with multiple peaks commonly complicate mapping and identification of catheter ablation targets in infarcts. METHODS: Retrospective analysis of catheter mapping data from eight patients with prior infarction was performed to evaluate multipotential electrograms at sites where pacing entrained VT. Potentials that were visible and not altered during pacing were defined as FFP. Potentials obscured by the pacing stimulus were designated possible LPs. The criteria for FFP were then assessed in a second cohort of five patients. RESULTS: At 32 of 39 (82%) sites with multiple potentials, entrainment identified one of the potentials as an FFP. Radiofrequency ablation, assessed at 15 sites, reduced the amplitude of LPs by 62%, without significant effect on FFP amplitude. At 56% of sites with multiple potentials, measuring the postpacing interval to an FFP would lead to erroneous classification of the site location relative to the reentry circuit. In prospective evaluation, double potentials were identified at 77 sites in infarcts; entrainment demonstrated an FFP at 66 (86%) sites. CONCLUSIONS: Far-field potentials are common during mapping in infarcts. Many can be distinguished from local potentials by entrainment, improving the accuracy of mapping.
No takes yet. Share an insight, caveat, or question.
Tung et al. (2003) conducted an observational in Ventricular tachycardia with prior infarction (n=13). Entrainment mapping was evaluated on Identification of far-field potentials at sites with multiple potentials. Entrainment mapping successfully identified far-field potentials at 86% of double potential sites in infarcts during prospective evaluation, improving the accuracy of ventricular tachycardia mapping.
Synapse has enriched 10 related papers. Consider them for comparative context: