Key result
The median recovery of spiked cTnI varied widely among assays (Vitros 29 ng/L to Roche 271 ng/L), with the Vitros assay having the highest percentage of outlier results (49%) exceeding the critical difference.
There is significant variability in the recovery of native cardiac troponin I across four routine assays, with outlier results likely driven by patient-specific interfering factors rather than microclots.
Clinicians should interpret cTnI results with assay-specific caution; leaves open standardization needs across platforms.
INTRODUCTION: This study aimed to examine the recovery of spiked human cardiac troponin I (cTnI) results measured by four routine assays, and investigate possible interference from microclots. MATERIALS AND METHODS: 457 consecutive samples with cTnI concentration below limit of quantitation (12 ng/L), declared by the Vitros TnI ES assay (reference assay), were measured on Beckman Coulter Accu TnI+3, Siemens TnI-Ultra and Roche TnI STAT assays. These samples were enriched with native full-length cTnI to a concentration of 100 ng/L and retested. A post-spiking result that exceeded the critical difference at a predefined probability of 0.0005 of the target concentration (the median post-spiking result for each individual assay) was considered as outlier. To determine whether microclots were a significant cause of critically discrepant outlier results, a separate 50 samples were centrifuged twice between two post-spiking measurements using the Vitros TnI ES assay. RESULTS: The median recovery of the enriched cTnI was highest with the Roche assay (271 ng/L) and lowest with the Vitros assay (29 ng/L). The Vitros assay had the highest percentage of results that exceeded the critical difference (49%), followed by the Siemens (38%), Roche (18%) and Beckman Coulter (7%) assays. None of the 50 additional samples produced a critically lower cTnI result after re-centrifugation. CONCLUSIONS: Our findings underscored the variability of cTnI assays in measuring native cTnI. The lack of cTnI results that became significantly lower after re-centrifugation suggested that microclots are unlikely to be a major cause of the outlier results.
No takes yet. Share an insight, caveat, or question.
Loh et al. (2016) studied Baseline cardiac troponin I below limit of quantitation (n=457). Spiked native human cardiac troponin I vs. Target concentration (median post-spiking result for each assay) was evaluated on Recovery of spiked cTnI and percentage of results exceeding the critical difference (outliers). The median recovery of spiked cTnI varied widely among assays (Vitros 29 ng/L to Roche 271 ng/L), with the Vitros assay having the highest percentage of outlier results (49%) exceeding the critical difference.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: