Answers are grounded in this paper and cite their sources.
Magnetic field drift is a significant factor that influences the long-term stability of the spin-exchange relaxation-free (SERF) comagnetometer. This work proposes a real-time closed-loop control method for the longitudinal compensation point of the comagnetometer. By developing a model of the transverse (x and y axes) oscillation magnetic field response of the K-Rb-21Ne comagnetometer, the relationship between the demodulation phase at a fixed driving frequency and the longitudinal (z-axis) remanence of the comagnetometer are provided in this paper. Based on this equation, a real-time closed-loop control system for the longitudinal axis compensation point is constructed, and the effectiveness of the closed-loop control method is verified by experiments. The experimental results show that the Allan variance at 100 s is reduced by 38.2% and the repeatability of the scale factor is improved by 83.3% in the closed-loop mode relative to the open-loop mode. The research work in this paper is crucial for advancing comagnetometer engineering applications.
No takes yet. Share an insight, caveat, or question.
Zhang et al. (2023) studied this question.