Key result
Heart rate monitors produced false-positive high heart rate artifacts in 45% of athletes and failed to detect single ectopic beats identified by Holter ECG in 15%.
Why the study?
Are heart rate monitors accurate tools for diagnosing exertion-induced arrhythmias in endurance athletes compared to Holter ECG?
Observational (n=142)
Are heart rate monitors accurate tools for diagnosing exertion-induced arrhythmias in endurance athletes compared to Holter ECG?
Heart rate monitors are prone to artifacts and are not reliable tools for diagnosing exercise-induced arrhythmias in asymptomatic athletes.
HRM-detected "arrhythmias" in asymptomatic athletes are typically artifacts; leaves open diagnostic utility pending ECG-confirmed prospective studies.
Millions of physically active individuals worldwide use heart rate monitors (HRMs) to control their exercise intensity. In many cases, the HRM indicates an unusually high heart rate (HR) or even arrhythmias during training. Unfortunately, studies assessing the reliability of these devices to help control HR disturbances during exercise do not exist. We examined 142 regularly training endurance runners and cyclists, aged 18-51 years, with unexplained HR abnormalities indicated by various HRMs to assess the utility of HRMs in diagnosing exertion-induced arrhythmias. Each athlete simultaneously wore a Holter electrocardiogram (ECG) recorder and an HRM during typical endurance training in which they had previously detected "arrhythmias" to verify the diagnosis. Average HRs during exercise were precisely recorded by all types of HRMs. No signs of arrhythmia were detected during exercise in approximately 39% of athletes, and concordant HRs were recorded by the HRMs and Holter ECG. HRMs indicated surprisingly high short-term HRs in 45% of athletes that were not detected by the Holter ECG and were artifacts. In 15% of athletes, single ventricular/supraventricular beats were detected by the Holter ECG but not by the HRM. We detected a serious tachyarrhythmia in the HRM and Holter ECG data with concomitant clinical symptoms in only one athlete, who was forced to cease exercising. We conclude that the HRM is not a suitable tool for monitoring heart arrhythmias in athletes and propose an algorithm to exclude the suspicion of exercise-induced arrhythmia detected by HRMs in asymptomatic, physically active individuals.
No takes yet. Share an insight, caveat, or question.
Gajda et al. (2017) conducted an observational in Unexplained heart rate abnormalities indicated by heart rate monitors (n=142). Heart rate monitors (HRMs) vs. Holter electrocardiogram (ECG) was evaluated on Utility of HRMs in diagnosing exertion-induced arrhythmias (concordance with Holter ECG). Heart rate monitors produced false-positive high heart rate artifacts in 45% of athletes and failed to detect single ectopic beats identified by Holter ECG in 15%.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: