Key result
Implantable pulse generator replacement successfully restored deep brain stimulation effect and normalized lead impedance in a patient with acute effect loss and high impedance without macroscopic fracture.
Case Report (n=1)
A staged approach starting with simple IPG replacement can be an effective option for acute DBS effect loss with high impedance when radiological findings are normal.
May support IPG replacement for unexplained high-impedance DBS failure; leaves open generalizability pending larger studies.
The number of deep brain stimulation (DBS) hardware complications has increased during the past decade. In cases of abnormally high lead impedance with no evidence of a macroscopic fracture, optimal treatment options have not yet been established. Here, we present the case of a 49-year-old woman with a 12-year history of Parkinson's disease who received bilateral subthalamic nucleus DBS in March 2006. The patient showed good control of parkinsonism until December 24, 2010, when she awoke with abrupt worsening of parkinsonian symptoms. At telemetric testing, lead impedances were found at >2,000 Ω in all four leads on the left side. Fracture of a lead or an extension wire was suspected. However, radiological screening and palpation revealed no macroscopic fracture. In June 2011, the implantable pulse generator (IPG) was changed under local anesthesia without any complications. Postoperatively, her parkinsonism immediately improved to the previous level, and the lead impedance readings by telemetry were also normalized. The disconnection of the neurostimulator connector block and the hybrid circuit board of the IPG was confirmed by destructive analysis. The present report illustrates that a staged approach that starts with simple IPG replacement can be an option for some cases of acute DBS effect loss with high impedance, when radiological findings are normal, thereby sparing the intact electrodes and extension wires.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2015) conducted a case report in Parkinson's disease with sudden loss of deep brain stimulation effect (n=1). Implantable pulse generator (IPG) replacement was evaluated on Improvement of parkinsonism and normalization of lead impedance. Implantable pulse generator replacement successfully restored deep brain stimulation effect and normalized lead impedance in a patient with acute effect loss and high impedance without macroscopic fracture.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: