Why the study?
Given the widespread use of hepatitis A serological assays in clinical laboratories, discrepancies among different immunoassay analyzers may have significant clinical implications.
Cross-Sectional (n=280)
No
Significant inter-assay variability exists in anti-HAV IgM detection across commercial automated immunoanalyzers, underscoring the need for improved harmonization.
Discrepancies in anti-HAV IgM across analyzers warrant interpretive caution in acute diagnosis; leaves open need for assay standardization.
BACKGROUND: Serological assays for hepatitis A virus (HAV) play a crucial role in diagnosing acute infections and monitoring disease transmission. Given their widespread use in clinical laboratories, discrepancies among different immunoassay analyzers may have significant clinical implications. This study aimed to assess the quantitative and qualitative agreement between anti-HAV total immunoglobulin (or IgG) and IgM results across four fully automated immunoassay systems. METHODS: A total of 280 and 223 clinical serum samples were tested for anti-HAV total immunoglobulin (or IgG) and IgM, respectively, using four immunoanalyzers: Vitros ECiQ (Ortho), Atellica IM 1600 (Siemens), Alinity i (Abbott), and Cobas e801 (Roche). Quantitative correlations and qualitative agreements were assessed, and cases with discordant anti-HAV IgM results were further investigated using available clinical data. RESULTS: While the total immunoglobulin (or IgG) assay demonstrated a strong correlation across all platforms, substantial discrepancies were observed in the IgM results, particularly between the Vitros ECiQ and the other three analyzers. Although the other three platforms yielded concordant results, the clinical review indicated that in 4 out of 6 cases (66.6%), the Vitros ECiQ results aligned more closely with the clinical presentations. CONCLUSIONS: This study highlights inter-assay variability in anti-HAV IgM detection and underscores the need for improved harmonization across platforms. Future studies incorporating a sufficient number of molecularly confirmed acute hepatitis A cases are warranted to clarify the causes of false results and minimize the potential clinical impact of inaccurate testing.
No takes yet. Share an insight, caveat, or question.
Song et al. (2025) conducted a cross-sectional in Hepatitis A virus (HAV) serological screening (n=280). Four fully automated immunoanalyzers (Vitros ECiQ, Atellica IM 1600, Alinity i, Cobas e801) vs. Consensus or Vitros ECiQ was evaluated on Qualitative agreement of anti-HAV total immunoglobulin (or IgG) and IgM results across four immunoanalyzers.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: