Key result
A fractured distal part of an Impella CP pump was successfully retrieved using a snare catheter under fluoroscopy after mechanical resistance hindered its removal.
Why the study?
Although Impella pump use has advanced, it is associated with risks, and intravascular device fracture during extraction is a rarely reported complication.
Case Report (n=1)
Device fracture is a rare but potential complication during the removal of a microaxial left ventricular pump, which can be safely managed using percutaneous snare retrieval under fluoroscopy.
Answers are grounded in this paper and cite their sources.
Impella fracture during extraction signals a rare procedural risk warranting vigilance; this case leaves open incidence and prevention strategies.
Background: The use of an Impella pump catheter has advanced substantially in the last few years due to the simple insertion procedure and smaller device size. However, its use is still associated with some risks and complications. Here, we report a device fracture as a rare complication that occurred during the device extraction a few days after the initial insertion. Case summary: A 74-year-old man with cardiogenic shock due to acute non-ST-segment elevation myocardial infarction presented to our hospital, and he was transferred to the cath lab for emergency percutaneous coronary intervention (PCI). An Impella CP pump was inserted without any complication prior to PCI. After successful PCI, the patient was transferred to the intensive care unit with device left for continued haemodynamic support. After 3 days, as the patient's condition remarkably improved, we tried to remove the device. However, a persistent mechanical resistance hindered the further catheter retraction; therefore, a decision was made to remove the catheter under fluoroscopy. Indeed, the fluoroscopy revealed a broken distal part of the pump at the level of the ascending aorta. The retained catheter tip was eventually snared with a snare catheter and removed without any complication. Discussion: An Impella microaxial pump may improve the overall outcome by providing haemodynamic support in critically ill patients. However, its application is not without complications. Intravascular device tip fracture, as demonstrated in this case report, is a rarely reported complication. The use of a snare catheter can be an option in retrieving a broken pump.
No takes yet. Share an insight, caveat, or question.
Mutallimov et al. (2022) conducted a case report in Cardiogenic shock due to acute non-ST-segment elevation myocardial infarction (n=1). Impella CP pump was evaluated on Device fracture and retrieval. A fractured distal part of an Impella CP pump was successfully retrieved using a snare catheter under fluoroscopy after mechanical resistance hindered its removal.
Synapse has enriched 10 related papers. Consider them for comparative context: