Why the study?
Does synchronized internal atrial defibrillation induce ventricular fibrillation in a patient with ventricular preexcitation?
Does synchronized internal atrial defibrillation induce ventricular fibrillation in a patient with ventricular preexcitation?
Synchronized internal atrial defibrillation can induce ventricular fibrillation in patients with ventricular preexcitation, which has important safety implications for the development of implantable atrial defibrillators.
Answers are grounded in this paper and cite their sources.
May induce VF in ventricular preexcitation; leaves open safety of internal atrial defibrillators.
This case describes ventricular proarrhythmia as a result of a synchronized internal atrial defibrillation shock in a 29-year-old man with Ebstein's anomaly referred for radiofrequency ablation of a right posterior accessory pathway. During the electrophysiologic study, atrial fibrillation was induced and 3/3 msec shocks of various strengths were delivered between two decapolar defibrillation catheters in the coronary sinus and right atrial appendage. A 2.0-J biphasic shock synchronized to an R wave after a short-long-short ventricular cycle length pattern with a preshock coupling interval of 245 msec induced ventricular fibrillation, which was externally defibrillated with 200 J. This observation has implications for the development of implantable atrial defibrillators.
No takes yet. Share an insight, caveat, or question.
Barold et al. (1997) studied this question.