The tested oxygen saturation sensor demonstrates high reliability and accuracy under various physiological conditions, supporting its potential use for chronic pacemaker control.
Answers are grounded in this paper and cite their sources.
Supports sensor development for adaptive pacing; leaves open human translation and outcome trials.
An oxygen saturation sensor, for the purpose of chronically controlling the heart rhythm produced by a pacemaker, should be specific to oxygen saturation and should be minimally affected by the harsh blood environment. For the sensor type we tested we found: (1) one sensor failure in 205.5 canine-months of chronic implantation (n = 11, range 4 to 50 months); (2) hematocrit-induced error of less than 5 percentage points of SvO2 over the range of 50% to 80% SvO2 and 15% to 45% hematocrit; (3) carboxyhemoglobin (HbCO)-induced error of less than 4 percentage points of SvO2 with HbCO up to 20%; (4) a fibrotic sheath-induced error of less than 3 percentage points of SvO2 in the range of 50% to 80% SvO2 due to fibrotic sheath thicknesses up to 0.22 mm; (5) no significant error induced by velocity variations local to the sensor; (6) no significant error due to temperature in the range of 30 degrees to 42 degrees C; and (7) that the sensor could be as close as 0.3mm to the ventricular wall and still only produce an error of 5% SvO2.