Key result
Steady-state and transient flow measurements for TAVR paravalvular leakage showed comparable trends but differing absolute values, with variation increasing up to 50% in imperfect annuli.
Why the study?
When establishing in silico models of paravalvular leakage to evaluate TAVR performance, the comparability between steady-state and transient PVL measurements as simulation inputs needed investigation.
Are steady-state flow measurements comparable to transient flow measurements for characterizing paravalvular leakage in TAVR models?
Are steady-state flow measurements comparable to transient flow measurements for characterizing paravalvular leakage in TAVR models?
Steady-state and transient flow measurements yield different absolute values for paravalvular leakage in TAVR models, complicating the extraction of generalized assumptions for in silico models.
Steady-state PVL measurements may diverge from transient conditions in TAVR models; leaves open accurate generalization for in silico simulations.
Severe aortic valve stenosis has been successfully treated with minimally invasive transcatheter aortic valve replacement (TAVR) for more than 20 years. The therapy is applied to an increasing cohort of patients. Nevertheless, paravalvular leakage (PVL) remains one of the main complications after TAVR implantation reducing safety and efficacy of the implant. To improve TAVR performance, virtual patient cohorts can be used to generate a significant amount of data with biologically varying conditions comparable to clinically realworld data. For this purpose, in silico models need to be established which are used to evaluate the possible occurrence of PVL. When establishing in silico models, the question between simplification to reduce computational effort and still valid information from the model must always be clarified. Thus, we investigated the comparability between steadystate and transient PVL measurements as an input parameter for simulations in virtual cohorts. A clinically established TAVR was implanted into different annulus models and tested in a steady-state back-flow test bench and in a pulse duplicator system. As a result, leakage rates were compared and assumptions for the in silico models were derived. Although the trend of PVL is comparable in steady-state and transient conditions, absolute values differ making it difficult to extract generalized assumptions. Especially when PVL increased, e.g. in a larger implantation diameter or in an imperfect annulus, the variation also increased up to 50% of the measured value.
No takes yet. Share an insight, caveat, or question.
Kaule et al. (2023) studied Paravalvular leakage after transcatheter aortic valve replacement. Steady-state flow measurements vs. Transient flow measurements was evaluated on Paravalvular leakage rates. Steady-state and transient flow measurements for TAVR paravalvular leakage showed comparable trends but differing absolute values, with variation increasing up to 50% in imperfect annuli.