Key result
The FIGO 2015 cardiotocography classification system yielded suboptimal interobserver agreement (κ=0.38), similar to NICE 2007 (κ=0.37) and NICE 2014 (κ=0.34) guidelines.
Why the study?
Does the FIGO 2015 cardiotocography classification system improve interobserver agreement compared to NICE 2007 and 2014 guidelines among labor ward staff?
Cross-Sectional (n=21)
No
Does the FIGO 2015 cardiotocography classification system improve interobserver agreement compared to NICE 2007 and 2014 guidelines among labor ward staff?
Absolute Event Rate: 0.38% vs 0.37%
The FIGO 2015 cardiotocography classification system offers favorable agreement scores and perceived ease of use compared to NICE guidelines, though overall interobserver agreement remains suboptimal.
No takes yet. Share an insight, caveat, or question.
Suboptimal interobserver agreement persists across CTG systems; supports preference for FIGO 2015 but leaves open impact on outcomes.
Bhatia et al. (2017) conducted a cross-sectional in Intrapartum cardiotocography (n=21). FIGO 2015 guidelines vs. NICE 2007 and NICE 2014 guidelines was evaluated on Interobserver agreement (κ values). The FIGO 2015 cardiotocography classification system yielded suboptimal interobserver agreement (κ=0.38), similar to NICE 2007 (κ=0.37) and NICE 2014 (κ=0.34) guidelines.