Basing on the fact that only short layers of a chromatographic column contribute to the separation in the interaction chromatography, 1 mm thick membranes from macroporous methacrylate polymer provided with functional groups were synthetized and used for protein separation. The chromatograms show that the separation is fully comparable with that experienced on a filled column but the advantage of a membrane is up to two orders of magnitude lower pressure during the process and very high loading reaching up to 40 g/m2. This recommends the high performance membrane chromatography also for large scale preparative separations.
No takes yet. Share an insight, caveat, or question.
Tennikova et al. (1990) studied this question.