Key result
Steerable sheath achieves ~45° deflection enabling successful LAAO after single transseptal puncture for M-TEER.
Why the study?
Atrial fibrillation frequently coexists with mitral disease in patients with high bleeding risk, generating interest in combined M-TEER and LAAO procedures requiring optimal coaxial sheath alignment.
Does the use of an Amplatzer steerable delivery sheath enable successful coaxial alignment and left atrial appendage occlusion following a supero-posterior transseptal puncture required for concurrent mitral transcatheter edge-to-edge repair?
Case Report (n=1)
No
Does the use of an Amplatzer steerable delivery sheath enable successful coaxial alignment and left atrial appendage occlusion following a supero-posterior transseptal puncture required for concurrent mitral transcatheter edge-to-edge repair?
Answers are grounded in this paper and cite their sources.
This case report demonstrates the feasibility of using a steerable delivery sheath to achieve coaxial alignment for left atrial appendage occlusion despite an unfavourable supero-posterior transseptal puncture required for concomitant mitral transcatheter edge-to-edge repair.
Background: Patients with atrial fibrillation (AF) have a five-fold increase in stroke events, and ∼90% of the thrombi develop in the left atrial appendage (LAA). Left atrial appendage occlusion (LAAO) has emerged as a safe and feasible alternative to oral anticoagulation (OAC) for stroke prevention in selected patients with non-valvular AF and contraindications to OAC. Atrial fibrillation is closely associated with mitral disease, and there is a growing interest in combined procedures. More than half of patients undergoing a mitral transcatheter edge-to-edge repair (M-TEER) suffer of AF and many have high or unacceptable bleeding risk. Case summary: We present a case of an 80-year-old woman suffering from paroxysmal AF, right carotid siphon aneurysm, and primary mitral regurgitation, with a high bleeding risk, who underwent a combined intervention of M-TEER and LAAO. Discussion: The combination of these two procedures is a logical step once the access to the left atrium is obtained with a transseptal puncture (TSP) and a transesophageal echocardiography (TEE) is in place to guide both procedures. The turning point in LAAO procedure is a correct TSP allowing coaxial alignment of the sheath with the LAA neck. Steerable delivery sheaths are promising dedicated tools, particularly in challenging anatomy or during combined procedures requiring different TSP positions.
No takes yet. Share an insight, caveat, or question.
Maiani et al. (2023) conducted a case report in Paroxysmal atrial fibrillation and severe primary mitral regurgitation with high bleeding risk (n=1). Combined mitral transcatheter edge-to-edge repair (MitraClip G4 XTW) and left atrial appendage occlusion (22 mm Amplatzer Amulet) using an Amplatzer steerable delivery sheath was evaluated on Procedural success of combined M-TEER and LAAO with coaxial alignment and absence of peri-device leak. An Amplatzer steerable delivery sheath achieved ~45° distal deflection and coaxial alignment for single-attempt 22 mm Amulet LAAO without peri-device leak after supero-posterior TSP for M-TEER.
Synapse has enriched 10 related papers. Consider them for comparative context: