Key result
Transcatheter aortic valve replacement with SAPIEN 3 Ultra was associated with a similar 1-year risk of death compared to SAPIEN 3 (5.8% vs. 5.5%; HR 1.07, 95% CI 0.81-1.90).
Why the study?
Literature comparing SAPIEN 3 Ultra with SAPIEN 3 was limited to short-term follow-up.
Does transcatheter aortic valve replacement with SAPIEN 3 Ultra improve 1-year clinical and echocardiographic outcomes compared to SAPIEN 3 in patients undergoing TAVR?
Cohort (n=1,155)
No
Does transcatheter aortic valve replacement with SAPIEN 3 Ultra improve 1-year clinical and echocardiographic outcomes compared to SAPIEN 3 in patients undergoing TAVR?
Hazard Ratio: 1.07 (95% CI 0.81–1.9)
Absolute Event Rate: 5.8% vs 5.5%
The SAPIEN 3 Ultra valve provides similar 1-year clinical outcomes to the SAPIEN 3 valve, with reduced mild paravalvular leak but a greater increase in transprosthetic gradient over time.
S3U shows similar 1-year clinical safety to S3 with less mild PVL but greater gradient rise; leaves open long-term hemodynamic implications for TAVR valve choice.
BACKGROUND: The newest-generation balloon-expandable valve, SAPIEN 3 Ultra (S3U), is expected to function well with the enhanced sealing skirt compared with SAPIEN 3 (S3). However, current literature on the comparison between these two valves is limited to short-term follow-ups. Therefore, we aimed to evaluate the 1-year outcomes and echocardiographic changes of S3U compared with S3. METHODS: We retrospectively identified patients who underwent transfemoral-transcatheter aortic valve replacement with S3U or S3 (20/23/26 mm) at our institution in 2018-2020. Outcomes were 1-year clinical events and echocardiographic parameters, and were compared between S3U and S3 after adjustment with inverse probability of treatment weighting. RESULTS: The S3U and S3 groups included 297 (25.7%) and 858 (74.3%) patients, respectively. There were no significant differences between the two groups in clinical events (death 5.8% vs. 5.5%, hazard ratio [HR] 1.07, 95% confidence interval [CI] 0.81-1.90; death or heart failure hospitalization 10.3% vs. 10.1%, HR 1.04, 95% CI 0.67-1.62). The S3U group had a lower prevalence of mild paravalvular leakage (PVL) (13.7% vs. 22.3%, p = 0.044), with similar moderate PVL (0.7% vs. 1.2%, p > 0.99). No significant differences were observed in aortic valve mean gradient and Doppler velocity index at 1 year. However, the S3U group had a larger increase in mean gradient from implantation to 1 year (median +4.70 vs. +1.63 mmHg, p < 0.001). CONCLUSIONS: S3U and S3 carried similarly favorable clinical event risks. Nonetheless, S3U was associated with less frequent mild PVL but a larger increase in transprosthetic gradient. Further studies are needed to determine the prognostic impact of these hemodynamic differences.
No takes yet. Share an insight, caveat, or question.
Isogai et al. (2022) conducted a cohort in Aortic valve replacement (n=1,155). SAPIEN 3 Ultra vs. SAPIEN 3 was evaluated on death (HR 1.07, 95% CI 0.81-1.90). Transcatheter aortic valve replacement with SAPIEN 3 Ultra was associated with a similar 1-year risk of death compared to SAPIEN 3 (5.8% vs. 5.5%; HR 1.07, 95% CI 0.81-1.90).