Key result
IABP malposition causes acute hepatic dysfunction that resolves promptly upon device removal.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
We describe a rare intraaortic balloon pump (IABP) vascular complication as a result of malpositioning of the IABP. A 61-year-old man with unstable angina underwent emergency coronary artery bypass grafting soon after the insertion of an IABP. Postoperative hemodynamics were stable, but acute hepatic dysfunction occurred on the second postoperative day. Doppler echography revealed the absence of hepatic arterial flow. The IABP was removed, and arterial flow was immediately restored. Thereafter, the hepatic function recovered rapidly. This is a rare case that demonstrates how IABP can cause mechanical abdominal arterial branch obstruction. Evaluations using Doppler echography are useful in detecting such IABP complications.
No takes yet. Share an insight, caveat, or question.
Hankei Shin (2000) conducted a case report in Acute ischemic hepatic failure secondary to intraaortic balloon pump malposition in unstable angina (n=1). Intraaortic balloon pump (IABP) malposition and subsequent removal was evaluated on Hepatic arterial flow and hepatic function recovery. Intraaortic balloon pump malposition caused acute hepatic dysfunction on postoperative day 2 in a 61-year-old man, which resolved rapidly upon device removal.