Why the study?
Does the Caudal Extradural Catheter Tray, Oxford Set allow for safe and successful caudal extradural catheterisation and analgesia in children?
Does the Caudal Extradural Catheter Tray, Oxford Set allow for safe and successful caudal extradural catheterisation and analgesia in children?
Answers are grounded in this paper and cite their sources.
The Caudal Extradural Catheter Tray, Oxford Set is highly successful and safe for performing extradural anaesthesia and analgesia in children.
Using a purpose designed set of equipment, the Caudal Extradural Catheter Tray, Oxford Set (B Braun Medical Ltd, Sheffield, UK) we have evaluated the ease of cannulation of the caudal space, and the subsequent success in threading extradural catheters and obtaining satisfactory analgesia via the caudal route. The set was evaluated in 91 children (age range: 1 day to 10 years). Cannulation of the caudal space was achieved in all patients, and catheterisation of the extradural space was successful in 96.7% of patients. Postoperative analgesia was satisfactory in 95% of children who had continuous extradural analgesia. There were no major complications or neurological sequelae associated with using the set. We found the Caudal Extradural Catheter Tray provides the necessary equipment to perform extradural anaesthesia and analgesia safely and successfully in children of a wide age range.
No takes yet. Share an insight, caveat, or question.
Bhandal et al. (2006) studied this question.