Key result
Epicardial admittance measurement successfully detected acute left ventricular dilation, demonstrating a significant correlation with endocardial crystal volume.
Why the study?
Does an epicardial admittance measurement system detect acute LV dilation in a porcine model?
Does an epicardial admittance measurement system detect acute LV dilation in a porcine model?
Answers are grounded in this paper and cite their sources.
Epicardial admittance measurement can accurately detect acute LV dilation by isolating the blood conductance signal from the myocardium, offering a potential future implantable warning system for heart failure.
There are two implanted heart failure warning systems incorporated into biventricular pacemakers/automatic implantable cardiac defibrillators and tested in clinical trials: right heart pressures, and lung conductance measurements. However, both warning systems postdate measures of the earliest indicator of impending heart failure: left ventricular (LV) volume. There are currently no proposed implanted technologies that can perform LV blood volume measurements in humans. We propose to solve this problem by incorporating an admittance measurement system onto currently deployed biventricular and automatic implantable cardiac defibrillator leads. This study will demonstrate that an admittance measurement system can detect LV blood conductance from the epicardial position, despite the current generating and sensing electrodes being in constant motion with the heart, and with dynamic removal of the myocardial component of the returning voltage signal. Specifically, in 11 pigs, it will be demonstrated that 1) a physiological LV blood conductance signal can be derived; 2) LV dilation in response to dose-response intravenous neosynephrine can be detected by blood conductance in a similar fashion to the standard of endocardial crystals when admittance is used, but not when only traditional conductance is used; 3) the physiological impact of acute left anterior descending coronary artery occlusion and resultant LV dilation can be detected by blood conductance, before the anticipated secondary rise in right ventricular systolic pressure; and 4) a pleural effusion simulated by placing saline outside the pericardium does not serve as a source of artifact for blood conductance measurements.
No takes yet. Share an insight, caveat, or question.
Porterfield et al. (2010) studied Acute left ventricular dilation (animal model) (n=15). Epicardial admittance measurement vs. Traditional conductance and 2D sonomicrometry was evaluated on Correlation between left ventricular blood conductance and endocardial crystal volume during neosynephrine-induced dilation (slope 1.16 mS/ml, p=0.017). Epicardial admittance measurement successfully detected acute left ventricular dilation, demonstrating a significant correlation with endocardial crystal volume.