Key result
Continuous 24-hour photoplethysmogram wrist-band monitoring correctly identified atrial fibrillation with 94.7% sensitivity and 96.9% specificity using a 30-minute detection time-frame.
Why the study?
Atrial fibrillation detection is challenging because it is often asymptomatic and paroxysmal.
Does continuous 24-h photoplethysmogram (PPG) monitoring accurately detect atrial fibrillation compared to 3-lead ambulatory ECG in hospitalized adults?
Observational (n=173)
Single-blind (investigators blinded to initial 12-lead resting ECG)
No
Does continuous 24-h photoplethysmogram (PPG) monitoring accurately detect atrial fibrillation compared to 3-lead ambulatory ECG in hospitalized adults?
Continuous wrist-band PPG monitoring provides reliable atrial fibrillation detection with high sensitivity and specificity, particularly using a 30-minute detection window, and is well-tolerated by patients.
Answers are grounded in this paper and cite their sources.
May facilitate ambulatory AF detection; hypothesis-generating and should not yet change practice.
Aim: Atrial fibrillation (AF) detection is challenging because it is often asymptomatic and paroxysmal. We evaluated continuous photoplethysmogram (PPG) for signal quality and detection of AF. Methods: PPGs were recorded using a wrist-band device in 173 patients (76 AF, 97 sinus rhythm, SR) for 24 h. Simultaneously recorded 3-lead ambulatory ECG served as control. The recordings were split into 10-, 20-, 30-, and 60-min time-frames. The sensitivity, specificity, and F1-score of AF detection were evaluated for each time-frame. AF alarms were generated to simulate continuous AF monitoring. Sensitivities, specificities, and positive predictive values (PPVs) of the alarms were evaluated. User experiences of PPG and ECG recordings were assessed. The study was registered in the Clinical Trials database (NCT03507335). Results: The quality of PPG signal was better during night-time than in daytime (67.3 ± 22.4% vs. 30.5 ± 19.4%, p < 0.001). The 30-min time-frame yielded the highest F1-score (0.9536), identifying AF correctly in 72/76 AF patients (sensitivity 94.7%), only 3/97 SR patients receiving a false AF diagnosis (specificity 96.9%). The sensitivity and PPV of the simulated AF alarms were 78.2 and 97.2% at night, and 49.3 and 97.0% during the daytime. 82% of patients were willing to use the device at home. Conclusion: PPG wrist-band provided reliable AF identification both during daytime and night-time. The PPG data’s quality was better at night. The positive user experience suggests that wearable PPG devices could be feasible for continuous rhythm monitoring.
No takes yet. Share an insight, caveat, or question.
Väliaho et al. (2022) conducted an observational in Atrial fibrillation (n=173). Continuous 24-h photoplethysmogram (PPG) monitoring vs. 3-lead ambulatory ECG was evaluated on Detection of atrial fibrillation (30-min time-frame) (F1-score 0.9536). Continuous 24-hour photoplethysmogram wrist-band monitoring correctly identified atrial fibrillation with 94.7% sensitivity and 96.9% specificity using a 30-minute detection time-frame.
Synapse has enriched 10 related papers. Consider them for comparative context: