Key result
Motion-induced artifacts in wrist PPGs reduced the sensitivity of life-threatening arrhythmia detectors by 45–48% for extreme bradycardia and by 13–32% for ventricular tachycardia.
Why the study?
A model for simulating motion-induced artifacts in wrist photoplethysmograms was needed to improve the realism of photoplethysmogram models.
Simulating realistic motion artifacts in wrist PPGs reveals significant drops in arrhythmia detection sensitivity, highlighting the need for artifact-robust detectors.
Answers are grounded in this paper and cite their sources.
Wrist PPG arrhythmia detectors warrant caution in ambulatory use; leaves open whether artifact mitigation can restore sensitivity in prospective validation.
A model for simulating motion-induced artifacts in the wrist photoplethysmogram (PPG) is proposed for the purpose to improve realism of PPG models. The database of day-long PPGs, acquired during cardiac rehabilitation, is used to extract artifact characteristics, which further serve as a basis for modeling artifacts in simulated PPGs with life-threatening arrhythmias. Depending on the recording, 14–49% of the PPG duration is corrupted by artifacts, mostly due to device displacement, forearm and hand motion. The artifact type influence on the performance of a life-threatening arrhythmia detector shows that the sensitivity drops by 45–48% for extreme bradycardia and by 13–32% for ventricular tachycardia. Poor contact causes 2–4 times more false alarms of ventricular tachycardia compared to the other artifact types under investigation. Simulation of realistic artifacts encountered in activities of daily living allows to comprehensively investigate arrhythmia detectors and understand the artifact types most negatively affecting detection performance. The proposed PPG artifact model is of importance for developing and testing artifact-robust arrhythmia detectors.
No takes yet. Share an insight, caveat, or question.
Paliakaitė et al. (2021) studied Life-threatening arrhythmias. Motion-induced artifacts in wrist photoplethysmogram (PPG) was evaluated on Sensitivity of life-threatening arrhythmia detector. Motion-induced artifacts in wrist PPGs reduced the sensitivity of life-threatening arrhythmia detectors by 45–48% for extreme bradycardia and by 13–32% for ventricular tachycardia.
Synapse has enriched 10 related papers. Consider them for comparative context: