Deflectable-sheath His-bundle pacing achieves stable thresholds in dextrocardia with complete AV block.
Case Report (n=1)
No
Combining a decapolar mapping catheter with a C304 deflectable delivery sheath enables safe and effective permanent His-bundle pacing in patients with dextrocardia and situs inversus totalis, where standard fixed-curve sheaths are oriented in the opposite direction.
Answers are grounded in this paper and cite their sources.
May enable His-bundle pacing in dextrocardia; leaves open generalizability beyond single-case experience.
Since Deshmukh and associates1 reported the successful implementation of His-bundle pacing (HBP) in 2000, previous studies have reported that HBP can decrease heart failure admission and preserve left ventricular ejection fraction compared to right ventricular (RV) pacing.2,3
No takes yet. Share an insight, caveat, or question.
Yoshimoto et al. (2019) conducted a case report in Complete atrioventricular block with dextrocardia and situs inversus totalis (n=1). Permanent His-bundle pacing using a combination of a 6F deflectable decapolar mapping catheter and an 8.5F C304 deflectable lead delivery sheath was evaluated on Successful permanent nonselective His-bundle pacing and pacing threshold stability at 5-month follow-up. Permanent His-bundle pacing using a deflectable sheath and mapping catheter achieved stable thresholds of 2.0 V at 0.4 ms at 5 months in a patient with dextrocardia and complete AVB.