Key result
Mobile phonocardiogram machine learning detects pathological newborn murmurs with ~92% accuracy.
Why the study?
Does a Support Vector Machine classifier applied to mobile phonocardiogram data accurately distinguish between healthy and pathological heart sounds in newborns?
Observational (n=116)
No
Does a Support Vector Machine classifier applied to mobile phonocardiogram data accurately distinguish between healthy and pathological heart sounds in newborns?
A mobile phonocardiogram system using a Support Vector Machine classifier can accurately distinguish between healthy and pathological heart sounds in newborns.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for mobile newborn murmur screening; leaves open prospective validation before any clinical adoption.
Phonocardiogram (PCG) monitoring on newborns is one of the most important and challenging tasks in the heart assessment in the early ages of life. In this paper, we present a novel approach for cardiac monitoring applied in PCG data. This basic system coupled with denoising, segmentation, cardiac cycle selection and classification of heart sound can be used widely for a large number of the data. This paper describes the problems and additional advantages of the PCG method including the possibility of recording heart sound at home, removing unwanted noises and data reduction on a mobile device, and an intelligent system to diagnose heart diseases on the cloud server. A wide range of physiological features from various analysis domains, including modeling, time/frequency domain analysis, an algorithm, etc., is proposed in order to extract features which will be considered as inputs for the classifier. In order to record the PCG data set from multiple subjects over one year, an electronic stethoscope was used for collecting data that was connected to a mobile device. In this study, we used different types of classifiers in order to distinguish between healthy and pathological heart sounds, and a comparison on the performances revealed that support vector machine (SVM) provides 92.2% accuracy and AUC = 0.98 in a time of 1.14 seconds for training, on a dataset of 116 samples.
No takes yet. Share an insight, caveat, or question.
Amiri et al. (2017) conducted an observational in Newborn heart murmurs (innocent vs. pathological) (n=116). Mobile phonocardiogram (PCG) diagnosis using Linear Support Vector Machine (SVM) vs. Echocardiography diagnosis by a cardiologist (and comparison against other machine learning classifiers) was evaluated on Diagnostic classification accuracy for distinguishing healthy (innocent) from pathological heart sounds. Linear support vector machine classification of mobile phonocardiogram recordings distinguished pathological from innocent newborn murmurs with 92.2% accuracy and an AUC of 0.98.