Key result
CoaguChek S portable INR monitoring correlates strongly with laboratory testing, achieving ~90% expanded agreement.
Observational (n=43)
Single-blind
No
Answers are grounded in this paper and cite their sources.
The accuracy and reproducibility of the CoaguChek S, and its clinical agreement with conventional laboratory international normalized ratio (INR) determination, were evaluated in an outpatient anticoagulation clinic setting. Forty-three patients provided 248 paired INR measurements for analysis. The paired results were highly correlated (r = 0.90). The mean coefficient of variation for the CoaguChek S for a random sample of 21 patients with three repeated tests each, was 4%. Clinical applicability was also measured by discrepant INR values, as defined in the literature by expanded and narrow agreement, and by INR values resulting in a different clinical decision by a blinded haematology registrar. Expanded agreement and narrow agreement between the two INR values occurred 90 and 88% of the time, respectively. The stricter criteria set down by the clinician resulted in 73% of paired results producing the same dosage decision. The CoaguChek S displayed good correlation with laboratory determination of INR and compared relatively well with expanded and narrow clinical agreement criteria.
No takes yet. Share an insight, caveat, or question.
Jackson et al. (2004) conducted an observational in Anticoagulation therapy monitoring (n=43). CoaguChek S portable INR monitor vs. Conventional laboratory INR determination was evaluated on Accuracy, reproducibility, and clinical agreement (expanded, narrow, and clinician dosage decision agreement) with laboratory INR (r = 0.90). CoaguChek S portable INR monitoring correlated strongly with laboratory INR, achieving 90% expanded agreement, 88% narrow agreement, and 73% identical dosage decisions.
Synapse has enriched 10 related papers. Consider them for comparative context: