Key points are not available for this paper at this time.
Design
Editorial
Answers are grounded in this paper and cite their sources.
Highlights the importance of quality control variables, such as thromboplastin choice and ISI assignation, for the viability and safety of the INR system in monitoring oral anticoagulation.
The introduction of the INR system for the monitoring of oral anticoagulant control represents a major advance, not only in terms of therapeutic efficacy but also in respect of patient safety. It is important to recognize that the system is only viable if careful attention is paid to the many important variables which contribute to its overall value. These include the choice of thromboplastin, assignation of ISI, determination of the mean normal prothrombin time and the method of end-point detection. Participation in independent external quality assessment schemes permits a unique opportunity for individual laboratories to identify, through laboratory performance analysis, problems relating to their own laboratory practice. Another major advantage, particularly of the larger schemes, is the identification of poor reagents and coagulometers and inconsistencies in their performance. One example of this is the recent identification of problems in respect of ISI assignment for thromboplastins used with coagulometers. This, in turn, provided the necessary stimulus to the potentially important development of calibrants for local ISI determination (21).
No takes yet. Share an insight, caveat, or question.
F E Preston (1995) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: