Why the study?
Can real-time, two-dimensional echocardiography diagnose pacemaker perforation in patients with suspected myocardial perforation?
Can real-time, two-dimensional echocardiography diagnose pacemaker perforation in patients with suspected myocardial perforation?
Real-time, two-dimensional echocardiography is a valuable tool for diagnosing and guiding the management of pacemaker perforation.
Answers are grounded in this paper and cite their sources.
May aid suspected pacemaker perforation diagnosis; leaves open routine adoption pending larger studies.
Five patients (four adults and one child) with clinically suspected myocardial perforation by temporary transvenous pacemakers were studied by real-time, two-dimensional echocardiography. In three patients, the catheters were visualized passing through the right ventricular apical wall with the tip located outside the cardiac border. In one patient the catheter perforated the atrioventricular septum and entered the left ventricle with the tip lodged against the posterior wall. In another patient, the catheter had partially penetrated the ventricular septum near the apex. Pericardial effusion was observed in two patients, in one of whom it was localized to the site of perforation. No patient had evidence of cardiac tamponade. In four patients, the catheters were withdrawn under echocardiographic visualization and the catheter tips could be seen moving from the abnormal locations back into the right-heart chambers. Perforation was verified at autopsy in two patients, including one in whom the catheter was not withdrawn. Real-time, two-dimensional echocardiography appears to be valuable in the diagnosis of pacemaker perforation.
No takes yet. Share an insight, caveat, or question.
Gondi et al. (1981) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: