Do cellular telephones cause electromagnetic interference with implantable cardioverter defibrillators?
Population
30 patients with implantable cardioverter defibrillators from various manufacturers.
Design
Cross-sectional
Authors
Loading...
No false detections seen despite telemetry noise; supports avoiding phones near programming head but leaves open risks with modern devices.
Do cellular telephones cause electromagnetic interference with implantable cardioverter defibrillators?
Cellular telephones do not cause false arrhythmia detections or under-sensing in ICDs, but they can disrupt telemetry and should be avoided during device interrogation and programming.
Occhetta et al. (1999) studied this question.