This is the authors' abstract. We don't add key points for this paper.
Multiple GaN nanowires produced by thermal chemical vapor deposition were employed as gas sensors for detection of hydrogen at concentrations from 200–1500 ppm in N2 at 300 K. Palladium coating of the wires improved the sensitivity by a factor of up to 11 at low ppm concentrations relative to uncoated controls. The GaN nanowires showed relative responses of ∼7.4% at 200 ppm and ∼9.1% at 1500 ppm H2 in N2 after a 10 min exposure. Upon removal of hydrogen from the measurement ambient, ∼90% of the initial GaN conductance was recovered within 2 min. Temperature dependent measurements showed a larger relative response and shorter response time at elevated temperature. The adsorption activation energy of the sensor was 2.2 kcal mol−1 at 3000 ppm H2 in N2. These sensors exhibit low power consumption (<0.6 mW) at 300 K.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Lim et al. (2008) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: