Why the study?
Blood flow adds variability to hepatic radiofrequency ablation outcomes, but few studies exist on its impact on temperature-controlled RFA.
Temperature-controlled radiofrequency ablation compensates for heat sink effects in larger vessels by increasing energy input, which can be accurately modeled in simulations.
Phantom and simulation results on perfusion effects in RFA leave open clinical translation and protocol optimization.
Purpose Blood flow is known to add variability to hepatic radiofrequency ablation (RFA) treatment outcomes. However, few studies exist on its impact on temperature‐controlled RFA. Hence, we investigate large‐scale blood flow effects on temperature‐controlled RFA in flow channel experiments and numerical simulations. Methods Ablation zones were induced in tissue‐mimicking, thermochromic phantoms with a single flow channel, using an RF generator with temperature‐controlled power delivery and a monopolar needle electrode. Channels were generated by molding the phantom around a removable rod. Channel radius and saline flow rate were varied to study the impact of flow on (i) the ablated cross‐sectional area, (ii) the delivered generator power, and (iii) the occurrence of directional effects on the thermal lesion. Finite volume simulations reproducing the experimental geometry, flow conditions, and generator power input were conducted and compared to the experimental ablation outcomes. Results Vessels of different channel radii affected the ablation outcome in different ways. For mm, the ablated area decreased with increasing flow rate while the energy input was hardly affected. For mm and mm, the energy input increased toward larger flow rates; for these radii, the ablated area decreased and increased toward larger flow rates, respectively, while still being reduced overall as compared to the reference experiment without flow. Directional effects, that is, local shrinking of the lesion upstream of the needle and an extension thereof downstream, were observed only for the smallest radius. The simulations qualitatively confirmed these observations. As compared to performing the simulations without flow, including flow effects in the simulations reduced the mean absolute error between experimental and simulated ablated areas from 0.23 to 0.12. Conclusion While the temperature control mechanism did not detect the heat sink effect in the case of the smallest channel radius, it counteracted the heat sink effect in the case of the larger channel radii with an increased energy input; this explains the increase in ablated area toward high flow rates (for mm). The experiments in a simple phantom setup, thus, contribute to a good understanding of the phenomenon and are suitable for model validation.
No takes yet. Share an insight, caveat, or question.
Nolte et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: