Answers are grounded in this paper and cite their sources.
Introduction: Thrombocytopenia is one of the most common pathologies encountered in laboratories. With reference to ISLH guidelines for smear review, a count of <100,000/μl should be reviewed with smear. Aims and Objectives: (1) To analyse the platelet scatterplot in cases of thrombocytopenia and decide on need for smear review. (2) To identify the cases of pseudo-thrombocytopenia through the platelet scatterplot pattern. Materials and Methods: This study was conducted from November 2023 to October 2024, at Neuberg Anand Reference Laboratory, Bengaluru. A total of 117,092 cases were processed in an automated analyser for platelet counts. Of these cases, 3118 cases of thrombocytopenia with defined inclusion criteria were considered. Results: A total of 1865 cases of moderate thrombocytopenia, 977 cases of severe thrombocytopenia and 276 cases of pseudo-thrombocytopenia were analysed. The platelet scatterplot was analysed in all the cases and were observed for pattern of events and subsequent distribution in the designated singlet/multiplet region. Conclusion: The automated haematology analysers are vital part of pathology laboratory and provide graphical data in the form of scatterplot. Platelet scatterplot is unique in cases of pseudo-thrombocytopenia. This study, being the first to analyse the platelet scatterplot with respect to reliability of platelet count and identify pre-analytical causes for thrombocytopenia, can be a valid primary tool in reviewing smear in cases of thrombocytopenia.
No takes yet. Share an insight, caveat, or question.
Venkatachala et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: